Using the Autoclave

An autoclave uses high pressure and temperature steam to effectively kill microorganisms or render a biohazardous material inactive. For effective sterilization, the materials/load must be saturated with this steam. Air pockets or insufficient steam supply will prevent effective decontamination.

If you have never operated the autoclave you will be using, contact an experienced user in your laboratory, the facilities manager or department safety contact for instruction on safe operation.

Autoclave Safety

Prevent injuries when using the autoclave by observing the following:

  • Never put volatile chemicals or solvents (alcohols, chloroform), corrosive chemicals (bleach, acetic acid, formalin, fixed tissues etc.), Sharps or radioactive materials in an autoclave.

  • Wear appropriate street clothing including closed-toe shoes and Personal Protective Equipment (PPE) including a lab coat, heat resistant gloves, and eye protection especially when unloading the autoclave.

  • Don't autoclave sealed containers as they pose an explosion risk.

  • Check the drain to make sure it isn't clogged.

  • Never open the door to the autoclave if there is water running out the bottom. Clogged steam lines, equipment malfunction or plugged drains may cause a buildup of scalding water.

  • Prevent steam burns and shattered glassware by making sure that the pressure in the autoclave chamber is zero and the temperature is at or below 121°C before opening the door at the end of a cycle. Make sure that no personnel, including you, are directly in front of the door.

  • Be especially careful with liquids due to superheating, a condition that occurs when liquids are at a temperature above their normal boiling point but do not appear to be boiling. In situations where personnel are in a hurry to remove flasks or bottles from the autoclave, the superheated liquids may boil out of their containers or explode.

Effective Sterilization

  • Place bags/containers in a tray rated for at least 135°C (stainless steel is recommended) for containment.

  • Open the bags a minimum of 2" so that the steam can get into the bags. If you don't want to remove the tape or twist tie, cut a large opening on the side of the bag. Consider adding 1 cup of water inside dry loads to generate additional steam.

  • Select the correct pre-programmed cycle if the unit has choices.

Autoclaving Biological Waste

Generally recommended parameters for biological waste are sterilization time of 60-120 minutes (excluding exhaust time) and 121°C at 15 psi.

Studies have shown that the processing time necessary to achieve decontamination of biological material depends on several loading factors. Load size, type of container, and moisture content all impact decontamination time. If the waste is packed too tightly or the autoclave is over packed the sterilization time may need to be increased to allow for steam penetration to the center of the load and reach 121°C. The cycle parameters may require different settings depending on the infectious materials you are trying to decontaminate. To assure that the material you are autoclaving is properly decontaminated/treated (killed or inactivated) add 10-15 minutes to the estimated sterilization time for over-kill. To verify if the materials are sterile/decontaminated, test the load with spore vials containing a biological indicator Geobacillus stearothermophilus. Contact DRS via e-mail for free spore vials and instructions for validation.