| Items | Biological Waste (Gravity Cycle) | Liquids(Liquid Cycle) | Dry Items (Gravity Cycle) | Glassware (Gravity Cycle) |
|---|---|---|---|---|
| Preparation | Open the bag >2", Place in tray, Place indicator if needed | Loosen caps or use a vented closure, Fill containers no more than 75% capacity | Fabrics Wrap; Instruments: Clean, dry, lay in pan | Dirty: Place in middle of the pan; Clean: wash, rinse, wrap |
| Placement in Autoclave | In the center | Upright in pan | Fabrics: Separated, on edge; Instruments: Flat | Dirty: In detergent and pan; Clean: On side or inverted |
| Temperature | 121°C | 121°C | 121°C | 121°C |
| Treatment Time in Minutes | 60-120 min. depending on load size and packing density | 22 min. for volumes <100mL; 40 min. for volumes >100mL | 30-60 min. | 30-60 min. |
| Exhaust Cycle | Slow exhaust | Slow exhaust | Fast exhaust and dry | Dirty: Slow exhaust; Clean: Fast/dry |
| Notes: | Avoid puncturing bags. Overbag and dispose of properly. | Hot bottles may explode. Let cool before moving. | Check reference for proper packaging methods | Glassware with cracks or deep scratches may crack |